LAM-562
Liquid Activity Monitor
Spectrometric measurement of the volumetric activity of gamma radionuclides in liquids in nuclear facilities.
Nuclear power plants
Waste management
Research centres
Key Features
- Off-line monitor with a measuring chamber
- Spectrometric measurement of volumetric activity of gamma emitting radionuclides in liquids
- Spectrum stabilization using a 40K source
- Compliant with IEC 60861 - Equipment for monitoring of radionuclides in liquid effluents and surface waters
Description
The LAM-562 monitor is designed for the continuous measurement of the volumetric activity of gamma-emitting radionuclides in liquids. It is primarily used in process circuits of nuclear power plants or research centres. It can also be used for controlled releases of liquids into the environment.
In a standard configuration, the monitor consists of the following parts:
- A frame with a shielded measuring chamber, through which the liquid sample flows (flow must be ensured by a pump)
- GD-52 Gamma Detector with a NaI(Tl) scintillator and a 40K stabilization source
- Digital spectrometric module DIM-15 with an MCA (1K/2K/4K/8K) for signal processing and a PLC for the archiving of measured spectra, measurement results, and events
- RDU-02 Radiation Display Unit (displaying up to 4 channels) or RDU-22 (displaying up to 16 channels)
Product design and specifications are subject to change without prior notice or obligation.